Vulnerability · CVE-2025-0683
Exposure of Private Personal Information to an Unauthorized Actor
Published Jan 20, 2026 · corrected/updated Jan 20, 2026 · Contec Health CMS8000 Patient Monitor · evidence confidence very high
In its default configuration, Contec Health CMS8000 Patient Monitor transmits plain-text patient data to a hard-coded public IP address when a patient is hooked up to the monitor. This could lead to a leakage of confidential patient data to any device with that IP address or an attacker in a machine-in-the-middle scenario.
Grounded briefings
Catalog interpretations for three audiences. Each brief cites ingested metadata and the original advisories below. They are not independent research and they do not invent exploit steps.
Executive
CVE-2025-0683 is a unknown public vulnerability record for Contec Health CMS8000 Patient Monitor. CISA lists it as known exploited. Published remediation guidance: Apply remediations or mitigations per vendor instructions or discontinue use of the product if remediation or mitigations are unavailable. This summary is a catalog interpretation of the ingested metadata and cites the original advisories below.
System administrator
Identify whether Contec Health CMS8000 Patient Monitor is in your estate, then compare installed versions with the affected ranges. Affected packages or ranges on record: Contec Health CMS8000 Patient Monitor. Published remediation guidance: Apply remediations or mitigations per vendor instructions or discontinue use of the product if remediation or mitigations are unavailable. Prefer the vendor or distribution advisory linked in Sources before changing production systems. This is a catalog interpretation, not a runbook.
Security analyst
Signals on file: no CVSS/EPSS yet. CISA lists it as known exploited. Sources contributing to this record: cve_org, vulncheck_kev. Public exploit references, when present, are outbound links only and are not proof of active exploitation. This is a catalog interpretation of stored metadata.
Exploitation and remediation
Exploitation evidence
Confirmed in CISA KEV since Jan 20, 2026. Listed in CISA KEV and VulnCheck KEV.
Patch and remediation
Apply remediations or mitigations per vendor instructions or discontinue use of the product if remediation or mitigations are unavailable.
Safe detection guidance
Compare installed versions with the affected ranges below and verify the advisory from the original publisher. This site stores metadata and outbound links only; it does not ship exploit code or proof-of-concept payloads.
Code
The project this record applies to. Exploit listings stay in References.
No public source repository is identified for this product.
Timeline
Detection, updates, and resolution for this record.
Jan 20, 2026 · VulnCheck KEV
Detected
In its default configuration, Contec Health CMS8000 Patient Monitor transmits plain-text patient data to a hard-coded public IP address when a patient is hooked up to the monitor. This could lead to a leakage of confidential patient data to any device with that IP address or an attacker in a machine-in-the-middle scenario.
Jan 20, 2026 · CISA KEV
Known exploited
Apply remediations or mitigations per vendor instructions or discontinue use of the product if remediation or mitigations are unavailable.
Affected
- vendor Contec Health CMS8000 Patient Monitorversion unspecified
Sources
- CVE.orgVulnerability recordCVE-2025-0683
Official CVE identifier
cve.org - VulnCheck KEVConfirmed exploitationCVE-2025-0683
In its default configuration, Contec Health CMS8000 Patient Monitor transmits plain-text patient data to a hard-coded public IP address when a patient is hooked up to the monitor. This could lead to a leakage of confidential patient data to any device with that IP address or an a
vulncheck.com