A vulnerability in the application software of multiple Radiometer products may allow remote code execution and unauthorized device management when specific internal conditions are

Published Dec 17, 2025 · catalog updated Dec 17, 2025 · Radiometer Medical Aps ABL90 FLEX and ABL90 FLEX PLUS Analyzers

Last checked Nov 2, 2026. A connected source was read within the last 7 days.

Check if you run it

No confirmed exploitation is on record for CVE-2025-14097. EPSS is 0.46%. Act only if a listed version of Radiometer Medical Aps ABL90 FLEX and ABL90 FLEX PLUS Analyzers is installed.

Gaps on this record

  • No vendor fix or required action is stored. Compare the linked advisory before you change production.
Severity
High7.2
Known exploitation
Not in CISA/VulnCheck KEV
EPSS
0.46%
Remediation
Version ranges on file
Affected products
vendor Radiometer Medical Aps ABL90 FLEX and ABL90 FLEX PLUS Analyzers Application software versions < 3.5MR11 with Windows 7, Windows XP operating systems

A vulnerability in the application software of multiple Radiometer products may allow remote code execution and unauthorized device management when specific internal conditions are met. Exploitation requires that a remote connection is established with additional information obtained through other means. The issue is caused by a weakness in the analyzer’s application software. Other related CVE's are CVE-2025-14095 & CVE-2025-14096. Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency. Required Configuration for Exposure: Affected

Are you affected?

Enter the version you run. The result uses only the ranges stored on this record.

The check returns affected, not affected, or incomplete. It does not guess a version that is missing from the record.

  • vendor Radiometer Medical Aps ABL90 FLEX and ABL90 FLEX PLUS Analyzers Application software versions < 3.5MR11 with Windows 7, Windows XP operating systemsApplication software versions < 3.5MR11 with Windows 7, Windows XP operating systems

How to fix CVE-2025-14097

  • Upgrade

    CVE-2025-14097 lists a fix boundary at 3.5MR11. Compare the installed build with that boundary before you change production.

  • Version check

    Seen in Application software versions < 3.5MR11 with Windows 7 · 1 other packages on the advisory

  • Advisory

    Open the cve_org record and follow its fix.

  • Official advisories

What to do

Three questions, each answered from this record only. Confirm the vendor advisory before you patch.

Should you patch?

No confirmed exploitation is on record for CVE-2025-14097. EPSS is 0.46%. Act only if a listed version of Radiometer Medical Aps ABL90 FLEX and ABL90 FLEX PLUS Analyzers is installed.

Which versions are affected?

Compare what you have installed with Radiometer Medical Aps ABL90 FLEX and ABL90 FLEX PLUS Analyzers Application software versions < 3.5MR11 with Windows 7, Windows XP operating systems Application software versions < 3.5MR11 with Windows 7, Windows XP operating systems. Compare the installed version with the linked vendor advisory before you change production.

What should an analyst verify?

No weakness class is stored. EPSS 0.46%. There is no CISA KEV listing. No Exploit-DB or Metasploit reference is linked. Use the source links for the original advisory.

Technology hubs

Code

The project this record applies to. Exploit listings stay in References.

No public source repository is identified for this product.

Timeline

Detection, updates, and resolution for this record.

  1. Dec 17, 2025 · EUVD

    Detected

    A vulnerability in the application software of multiple Radiometer products may allow remote code execution and unauthorized device management when specific internal conditions are met. Exploitation requires that a remote connection is established with additional information obtained through other means. The issue is caused by a weakness in the analyzer’s application software. Other related CVE's

  2. Dec 17, 2025 · EUVD

    Updated

    A vulnerability in the application software of multiple Radiometer products may allow remote code execution and unauthorized device management when specific internal conditions are met. Exploitation requires that a remote connection is established with additional information obtained through other means. The issue is caused by a weakness in the analyzer’s application software. Other related CVE's

Sources

  • CVE.orgVulnerability recordCVE-2025-14097

    Official CVE identifier

    cve.org
  • FIRST EPSSExploitation probabilityCVE-2025-14097

    30-day exploitation probability 0.46%

    first.org
  • EUVDVulnerability recordEUVD-2025-203895

    A vulnerability in the application software of multiple Radiometer products may allow remote code execution and unauthorized device management when specific internal conditions are met. Exploitation requires that a remote connection is established with additional information obta

    euvd.enisa.europa.eu

References

Freshness

Last verified in this catalog Nov 2, 2026. Significant claims (KEV, EPSS, CVSS, affected ranges) come from the source rows above — open the original publisher for the authoritative advisory.

  • FIRST EPSSNov 2, 2026
  • CVE.orgNov 2, 2026
  • EUVDNov 1, 2026