Vulnerabilities

Vulnerability · CVE-2026-7251

Eppendorf BioFlo 320 is vulnerable due to VNC server using a hard-coded password.

Published May 26, 2026 · catalog updated Jul 24, 2026 · Eppendorf BioFlo 320 is

Check if you run it

No confirmed exploitation is on record for CVE-2026-7251. EPSS is 0.74%. Act only if a listed version of Eppendorf BioFlo 320 is is installed.

Gaps on this record

  • The stored product name is a package slug. The name on this page is taken from the title (Eppendorf BioFlo 320 is).
  • No version range is stored. A customer cannot check an installed version against this record.
  • No vendor fix or required action is stored. Compare the linked advisory before you change production.
Severity
Critical9.3
Known exploitation
Not in CISA/VulnCheck KEV
EPSS
0.74%
Remediation
Not established
Affected products
See sources for product scope

Eppendorf BioFlo 320 is vulnerable due to VNC server using a hard-coded password. If a remote attacker knows the network address of any BioFlo 320 model with remote access enabled, they can gain full control of the user interface by using this password. Once connected, the attacker would have full access to all control panel features for the BioFlo 320. VNC traffic is not encrypted.

No affected version ranges are stored yet. Use the vendor advisory before you decide this record does not apply.

How to fix CVE-2026-7251

What to do

Three questions, each answered from this record only. Confirm the vendor advisory before you patch.

Should you patch?

No confirmed exploitation is on record for CVE-2026-7251. EPSS is 0.74%. Act only if a listed version of Eppendorf BioFlo 320 is is installed.

Which versions are affected?

Affected version ranges are not on this record. Compare the installed version with the linked vendor advisory before you change production.

What should an analyst verify?

CWE-259. EPSS 0.74%. There is no CISA KEV listing. No Exploit-DB or Metasploit reference is linked. Use the source links for the original advisory.

Code

The project this record applies to. Exploit listings stay in References.

Timeline

Detection, updates, and resolution for this record.

  1. May 26, 2026 · NIST NVD

    Detected

    Eppendorf BioFlo 320 is vulnerable due to VNC server using a hard-coded password. If a remote attacker knows the network address of any BioFlo 320 model with remote access enabled, they can gain full control of the user interface by using this password. Once connected, the attacker would have full access to all control panel features for the BioFlo 320. VNC traffic is not encrypted.

  2. Jul 24, 2026 · NIST NVD

    Updated

    Eppendorf BioFlo 320 is vulnerable due to VNC server using a hard-coded password. If a remote attacker knows the network address of any BioFlo 320 model with remote access enabled, they can gain full control of the user interface by using this password. Once connected, the attacker would have full access to all control panel features for the BioFlo 320. VNC traffic is not encrypted.

Sources

  • CVE.orgVulnerability recordCVE-2026-7251

    Official CVE identifier

    cve.org
  • FIRST EPSSExploitation probabilityCVE-2026-7251

    30-day exploitation probability 0.7%

    first.org
  • NIST NVDVulnerability recordCVE-2026-7251

    Eppendorf BioFlo 320 is vulnerable due to VNC server using a hard-coded password. If a remote attacker knows the network address of any BioFlo 320 model with remote access enabled, they can gain full control of the user interface by using this password. Once connected, the attack

    nvd.nist.gov

References

Freshness

Last verified in this catalog Oct 18, 2026. Significant claims (KEV, EPSS, CVSS, affected ranges) come from the source rows above — open the original publisher for the authoritative advisory.